Inter-Laboratory Comparability of Clinical Chemistry Testing: A New Perspective

Authors

  • Perez Quartey Department of Medical Laboratory Technology, Garden City University College, Kenyase, Ghana
  • Ruth Anokye Department of Medical Laboratory Technology, Garden City University College, Kenyase, Ghana
  • Michael Frimpong Department of Medical Laboratory Technology, Garden City University College, Kenyase, Ghana

Keywords:

Inter-laboratory, Mentor-adept, External Quality Assurance, z-score, Precision Index, Total Analytical Error

Abstract

Background/Aim: Existing external quality assurance proficiency testing may not reflect routine ‘real-world’ daily testing proficiency. The overall aim of the study was to conduct an inter-laboratory comparability testing of some routinely measured clinical chemistry analytes by selected private medical laboratories using the mentor-adept approach. Methodology: Aliquots of freshly separated plasma from a single individual (1ml) each were sent to each of the selected labs and a Teaching hospital for same-day testing. All samples were sent as ‘‘blinded-samples’’ (labelled with anonymous names, ages and attached lab request forms with clinical diagnoses) so that they would be tested as real patient samples. Results: All the labs met the acceptability criteria range for both the z-score and Precision Index. However, many of the labs had their total analytical errors for the tests outside the allowable total error ranges with both European and CLIA recommendations. Conclusion: There are relative similarities in the z-score (inter-laboratory bias) and Precision Index (inter-laboratory precision) among the labs. However, many of the labs did not meet recommended analytical goals for total analytical errors on individual samples run in a day. In the light of the findings, it is highly recommended that though laboratories should be aware of conduction of periodic external quality assurance exercises, such exercises should be done using “blinded-samples” as utilized in this current study without prior notification of the day and time of testing.

References

. Theodorsson E Magnusson B, Leito I. Bias in clinical chemistry. Bioanalysis 2014; 6(21), 2855–2875.

. ISO, IEC 17025 (2005) General requirements for the competence of testing and calibration laboratories. International Organization for Standardization/International Electrotechnical Commission, Geneva. https://www.iso.org/standard/39883.html. Accessed Feb 2019.

. EA-4/18 (2010) Guidance on the level and frequency of proficiency testing participation. http://www.european-accreditation.org/publication/ea-4-18-inf. EA-4/18: 2010. Accessed Feb 2019.

. Karkalousos P, Evangelopoulos A (2011). Quality Control in Clinical Laboratories, Applications and Experiences of Quality Control, Prof. Ognyan Ivanov (Ed.), ISBN: 978-953-307-236-4. http://www.intechopen.com/books/applications-and-experiences-of-quality-control/quality-control-inclinical-laboratories. Accessed Feb 2019.

. Szewczak E, Bondarzewski A. Is the assessment of interlaboratory comparison results for a small number of tests and limited number of participants reliable and rational?

Accred Qual Assur 2016; 21:91–100.

. ISO 13528 (2005) Statistical methods for use In proficiency testing by interlaboratory comparisons. International Organization for Standardization, Geneva. https://www.iso.org/standard/35664.html. Accessed Feb 2019.

. Hyltoft Peterson P, Ricos C, Stockl D, Libeer JC, Baadenhuijsen H, Fraser C, Thienpont L. ‘‘Proposed guidelines for the internal quality control of analytical results in the medical laboratory.’’ Eur J Clin Chem Clin Biochem 1996; 34:983-999.

. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, Minchinela J, Perich C, Simon M (2014) Desirable specifications for total error, imprecision, and bias, derived from intra- and inter-individual biologic variation. http://www.westgard.com/biodatabase1.htm. Accessed Feb 2019.

. Sciacovelli L, Secchiero S, Zardo L, Plebani M. The role of the External Quality Assessment. Biochemia Medica. 2010; 20(2):160–4.

. Kettelhut MM, Chiodini PL, Edwards H, Moody A. External quality assessment programmes raise standards: evidence from the UKNEQAS parasitology subschemes. J Clin Pathol. 2003; 56(12):927–32.

. ISO 13528 (2015) Statistical methods for use in proficiency testing by interlaboratory comparisons. International Organization for Standardization, Geneva. https://www.iso.org/standard/56125.html. Accessed Feb 2019.

. 2019 CLIA proposed changes to PT acceptable limits. https://www. westgard.com/2019-clia-changes.htm. Accessed April 2019.

. Fearn T, Fisher S, Thompson M, Ellison SLR. “A decision theory approach to fitness-for purpose in analytical measurement”, Analyst 2002; 127:818-824.

. Kristensen GB, Meijer P. Interpretation of EQA results and EQA-based trouble shooting. Biochem Med 2017 27: 49-62.

. Sareen R. Illuminating Z-Score in External Quality Assessment for Medical Laboratory. Health Care Current Review 2018, 6:3.

. ISO/IEC 17043:2010. Conformity assessment – General requirements for proficiency testing. Geneva: International Organization for Standardization; 2010.

. ISO 15189:2013. Medical laboratories – Requirements for quality and competence. International Organization for Standardization; 2013.

Downloads

Published

2019-11-16

How to Cite

Quartey, P., Anokye, R., & Frimpong, M. (2019). Inter-Laboratory Comparability of Clinical Chemistry Testing: A New Perspective. International Journal of Sciences: Basic and Applied Research (IJSBAR), 48(5), 196–207. Retrieved from https://gssrr.org/index.php/JournalOfBasicAndApplied/article/view/10475

Issue

Section

Articles